GF 8850电导率表说明书
Calibration Procedure
This instrument has been electronically calibrated at the factory.
Procedure A verifies the accuracy and linearity of the instrument by simulating temperature and conductivity values with precision
(±0.1%) fixed resistors.
Procedure B is a wet calibration. This procedure uses the sensor input and NIST traceable test solutions. When done correctly,
this procedure offers the most accurate system calibration.
page 8 of 12 SIGNET 8850-1 Conductivity/Resistivity Transmitter