GF 8850电导率表说明书
3.3Open Collector Output
The Open Collector output can be used as a switch that respondswhen the process value moves above or below a setpoint, or itcan be used to generate a pulse at a rate proportional to theprocess value.
Low:
Output triggers when process variable is less than the setpoint.The output will relax when the process variable moves above thesetpoint plus the hysteresis value.
High:
Output triggers when process variable is greater than the
setpoint. The output will relax when the process variable movesbelow the setpoint plus the hysteresis value.
Proportional Pulsing
The Open Collector output will generate a 100 mS pulse at therate defined by settings in the CALIBRATE menu.
In the example below:
The output will be 0 pulses/min. when value is less than 5. The output will be 50 pulses/min. when value is 7.5. The output will be 100 when value is greater than 10.The starting point, endpoint and maximum pulse rate areselectable in the CALIBRATE menu.
Hysteresis
Low Setpoint
Output activeOutput inactive
Starting point
High Setpoint
endpoint
VIEW menu
During normal operation, ProcessPro displays the VIEW menu.
When editing the CALIBRATE or OPTIONS menus, ProcessPro will return to the VIEW menu if no activityoccurs for 10 minutes.
To select a VIEW display, press the UP or DOWN arrow keys. The selections will scroll in a continuousloop.
Changing the VIEW display does not interrupt system operations.No key code is necessary to change display selection.Output settings cannot be edited from the VIEW menu.
View Menu
Display
Description
Monitor the Conductivity and Temperature values from the sensor.This is the permanent display.
123.45 µS/cm+67.89 ºC
The VIEW displays below are temporary. The permanent display will return after 10 minutes
Loop Output:13.75 mALast CAL:06-30-01
Monitor the Loop Current output.Check the Last Calibration date.
page 4 of 12 SIGNET
8850-1 Conductivity/Resistivity Transmitter